MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-05-26 for ARCHITECT TACROLIMUS 01L77-25 manufactured by Abbott Laboratories.
[76191749]
An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. An evaluation is in-process.
Patient Sequence No: 1, Text Type: N, H10
[76191750]
The customer observed one patient with a falsely elevated tacrolimus result while using the architect tacrolimus assay. The results were generated during treatment for a patient undergoing acute rejection for a heart transplant (received (b)(6) 2016). The customer indicated that 2 consecutive values which had been drawn on (b)(6) 2017 showed a falsely elevated, as follows: on (b)(6) 2017: 11. 2 mg/dl. On (b)(6) 2017: 44. 2 mg/dl. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1415939-2017-00118 |
| MDR Report Key | 6594247 |
| Report Source | HEALTH PROFESSIONAL |
| Date Received | 2017-05-26 |
| Date of Report | 2017-07-20 |
| Date of Event | 2017-05-03 |
| Date Mfgr Received | 2017-06-20 |
| Device Manufacturer Date | 2017-02-01 |
| Date Added to Maude | 2017-05-26 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | NOEMI ROMERO-KONDOS, RN BSN |
| Manufacturer Street | 100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3 |
| Manufacturer City | ABBOTT PARK IL 600643537 |
| Manufacturer Country | US |
| Manufacturer Postal | 600643537 |
| Manufacturer Phone | 224667-512 |
| Manufacturer G1 | ABBOTT LABORATORIES |
| Manufacturer Street | 100 ABBOTT PARK ROAD |
| Manufacturer City | ABBOTT PARK IL 600643500 |
| Manufacturer Country | US |
| Manufacturer Postal Code | 600643500 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | ARCHITECT TACROLIMUS |
| Generic Name | TACROLIMUS |
| Product Code | MLM |
| Date Received | 2017-05-26 |
| Catalog Number | 01L77-25 |
| Lot Number | 72006M800 |
| Device Expiration Date | 2018-06-01 |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | Y |
| Device Age | DA |
| Device Eval'ed by Mfgr | Y |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | ABBOTT LABORATORIES |
| Manufacturer Address | 100 ABBOTT PARK ROAD ABBOTT PARK IL 600643500 US 600643500 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2017-05-26 |