SMARTDOSE GAS CONSERVER, 310G CTOX-MN02

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,consum report with the FDA on 2017-05-31 for SMARTDOSE GAS CONSERVER, 310G CTOX-MN02 manufactured by Inovo Inc.

Event Text Entries

[76374787] The device is used to provide supplemental oxygen to patient with compromised respiratory function. It is not intended to be used as a life sustaining device. It was stated that the device was in conserve mode, which means the device producing a bolus of supplemental oxygen as the patient inhales. If the patient's breathing was shallow after the syncopal episode, there may not have been enough negative pressure for the device to deliver the oxygen at conserve mode. Once the device oxygen flow function was switched to continuous flow a steady flow of oxygen was delivered to the patient. It was stated that the patient was easily aroused and regained consciousness spontaneously with no medical intervention. These factors conclude that the patient did not suffer a loss of consciousness, but rather a syncopal episode. Syncopal episodes are a temporary loss of consciousness and posture, described as "fainting" or "passing out". It usually related to temporary insufficient blood flow to the brain. It most often occurs when the blood pressure is too low (hypotension) and the heart does not pump a normal supply of oxygen to the brain. If the patient had lost consciousness due to decreased oxygen to the brain (hypoxia), the patient would have needed more aggressive respiratory treatment than just simply switching the device to continuous flow mode. Syncope can be caused by neurological disorder, orthostatic hypotension or a side effect of medications. It can be concluded that the patient's temporary loss of consciousness was caused by some factor other than the amount of oxygen given or not given by the device. With the information provided, there is insufficient evidence to determine a causal relationship between the alleged in the complaint and the patient's syncopal episode; however, because a patient reportedly had a syncopal episode associated with an inovo device, the event was reported to the fda. The device has not yet been received for testing. Inovo, inc. Will submit a supplemental report if the device is received and after inovo, inc. Service/repair facility concludes its investigation. Note that this mdr was previously submitted on 02/03/2017 and is being re-submitted because it was not fully processed or received by fda due to an error.
Patient Sequence No: 1, Text Type: N, H10


[76374788] It was reported that the conserver was not producing sufficient oxygen. Additional information obtained from the delivery driver who spoke with the patient's caregiver is as follows: the caregiver stated that she walked into the room and found the patient had a syncopal episode. The caregiver stated that all the lights on the conserver were on and the device was in the conserve mode. She stated that she changed the device to continous flow and the patient regained consciousness. There was no reported device malfunction and it could not be determined if the patient regained consciousness because the device was switched to continous flow or if the symptoms resolved spontaneously. There was no additional reported medical interventions performed and the patient did not seek medical attention for the issue described. It was noted by the delivery driver that the tag with the serial number had faded as if a chemical cleaning agent was used on the device. No additional information provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1062191-2017-00007
MDR Report Key6603063
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2017-05-31
Date of Report2017-01-17
Date Mfgr Received2017-01-09
Device Manufacturer Date2016-05-04
Date Added to Maude2017-05-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRESPIRATORY THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMARITA BRUST
Manufacturer Street401 LEONARD BLVD NORTH
Manufacturer CityLEHIGH ACRES FL 33971
Manufacturer CountryUS
Manufacturer Postal33971
Manufacturer Phone5166694910
Manufacturer G1INOVO INC
Manufacturer Street401 LEONARD BLVD NORTH
Manufacturer CityLEHIGH ACRES FL 33971
Manufacturer CountryUS
Manufacturer Postal Code33971
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSMARTDOSE GAS CONSERVER, 310G
Generic NameCONSERVER
Product CodeNFB
Date Received2017-05-31
Model NumberCTOX-MN02
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINOVO INC
Manufacturer Address401 LEONARD BLVD NORTH LEHIGH ACRES FL 33971 US 33971


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-05-31

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.