COBAS 4000 C (311) STAND ALONE SYSTEM C311 04826876001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-06-02 for COBAS 4000 C (311) STAND ALONE SYSTEM C311 04826876001 manufactured by Roche Diagnostics.

Event Text Entries

[76666050] This event occurred in (b)(6). (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[76666051] The customer complained of an erroneous low result for 1 patient sample tested for crpl3 c-reactive protein gen. 3 (crpl3) on a cobas 4000 c (311) stand alone system. The initial crpl3 result was 0. 00 mg/l with a data flag. The result was reported outside of the laboratory as <1 mg/l. The nurse noticed that this result did not correspond to the patient? S historical results and requested the sample be repeated. The repeat result approximately an hour later was 237. 34 mg/l. The patient was not treated based on the results. No adverse event occurred. The patient has been discharged from the hospital. The crpl3 reagent lot number and expiration date were not provided. The customer said only one sample was affected, they had no issues prior to the event, and as of (b)(6) 2017 had had no further issues. The field service representative (fsr) visited the customer site and noted that water rinse levels were low. These were adjusted. The sample probe was replaced and adjusted. Gear head pump pressure was increased. Quality control results were acceptable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01135
MDR Report Key6608557
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-06-02
Date of Report2017-06-02
Date of Event2017-05-02
Date Mfgr Received2017-05-12
Date Added to Maude2017-06-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 4000 C (311) STAND ALONE SYSTEM
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeDCN
Date Received2017-06-02
Model NumberC311
Catalog Number04826876001
Lot NumberNA
ID NumberNA
OperatorBIOMEDICAL ENGINEER
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 4000 C (311) STAND ALONE SYSTEM
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-06-02
Model NumberC311
Catalog Number04826876001
Lot NumberNA
ID NumberNA
OperatorBIOMEDICAL ENGINEER
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-02

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.