970 SE-10 HOSPITAL SMART MONITOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2006-01-06 for 970 SE-10 HOSPITAL SMART MONITOR manufactured by Respironics, Inc..

Event Text Entries

[16142895] Information was received from the repair technician that on evaluation, this device was incapable of sounding the optional remote alarm to alert the caregiver of an event. There was no report of patient harm or injury with this device. The primary alarm on the device was found to operate correctly. Per complaint clarification from the customer the remote alarm option is not utilized by their facility, however the inability of the device to sound the optional alarm could cause or contribute to harm or injury to the intended user. The circuit board was replaced to correct the problem.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2518422-2006-00001
MDR Report Key661027
Report Source05,07
Date Received2006-01-06
Date of Report2005-12-17
Date of Event2005-12-17
Date Mfgr Received2005-12-17
Device Manufacturer Date2002-02-01
Date Added to Maude2006-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLOU KINNEY
Manufacturer Street1001 MURRY RIDGE LANE
Manufacturer CityMURRYSVILLE PA 15866
Manufacturer CountryUS
Manufacturer Postal15866
Manufacturer Phone7243874292
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand Name970 SE-10 HOSPITAL SMART MONITOR
Generic NameINFANT APNEA MONITOR
Product CodeBZQ
Date Received2006-01-06
Returned To Mfg2005-12-13
Model Number970 SE-10
Catalog Number970 SE-10
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key650444
ManufacturerRESPIRONICS, INC.
Manufacturer Address1001 MURRY RIDGE LN. MURRYSVILLE PA 156688550 US
Baseline Brand Name970 SE-10 HOSPITAL SMART MONITOR
Baseline Generic NameINFANT APNEA MONITOR
Baseline Model No970 SE-10
Baseline Catalog No970 SE-10
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2006-01-06

© 2026 FDA.report