75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CML-75LN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-05 for 75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CML-75LN manufactured by Charter Medical, Ltd..

Event Text Entries

[76836033] The cml-75ln is a cryogenic storage container consisting of a 75-ml bag with a fill tube and 2 spike ports. A pre-printed bag label is molded into the side of the bag during manufacturing. The label is intended to provide a means for recording contents information and remain permanently attached to the bag throughout the cell preservation and transfer/transfusion processes. A single used bag was received for investigation. Adhesive from what appeared to be a user applied label was present on the bag surface. The cml-75ln label, molded into the side of the bag, had been removed, presumably by the user. An irregular film tear, approximately 4-inches in length, was observed on the top portion of the bag. There is a possibility that a label attached to the bag surface could interfere with the flexibility of the film during cryogenic freezing and thawing processes. There are other causes that could potentially result in a tear: - overfilling the bag. An overfilled bag could create resistance during insertion and removal of the bag from the metal cassette. - excessive air left inside the bag after filling, resulting in rapid expansion during thawing. - moisture on the exterior surface of the bag or cassette interior when the bag is inserted into the metal cassette. Moisture may cause the bag film to freeze to the interior cassette surface, thus potentially resulting in damage when the frozen bag is removed from the cassette. - although there was no evidence, there is always the possibility that an undetected material issue could contribute to a tear of this nature. The product instruction for use sheet is provided with each cml-75ln device. Review of the ifu version provided with lot 145438 has the following precautions: - "do not write on bag surface, back side of label or adhere labels to bag surface. " - "after freezing, do not handle excessively. Port tubes and film are fragile in the frozen state and breakage may occur. Handle with care. " - "do not overfill. " - "remove as much air as possible from the container. " - "ensure bag exterior and protective freezing cassettes are dry prior to initiating freezing protocol. Moisture on the exterior of the bag or on the cassette could cause adherence of the bag to the cassette resulting in difficulty of bag removal. "
Patient Sequence No: 1, Text Type: N, H10


[76836034] On (b)(6) 2017, a cml-75ln device containing frozen cell product was being prepared for transfusion to a patient. While the cml-75ln device was undergoing the thawing process, a loud crack (pop) was heard and a bag break was noticed. The product was not in contact with a patient during the thawing process. Prophylactic antibiotics were administered to the patient prior to the transfusion process. The transfusion facility salvaged the cell material from the cml-75ln device, and infused an unknown amount of the cell material into the patient. A precautionary antibiotic was administered to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2017-00005
MDR Report Key6612734
Date Received2017-06-05
Date of Report2017-06-05
Date of Event2017-05-04
Date Mfgr Received2017-05-05
Device Manufacturer Date2016-05-03
Date Added to Maude2017-06-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA HUGHES
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Generic NameCELL FREEZE CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2017-06-05
Returned To Mfg2017-05-16
Model NumberCML-75LN
Catalog NumberCML-75LN
Lot Number145438
Device Expiration Date2019-04-01
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.