BARD BRACHYSTAR BRACHYTHERAPY APPLICATOR NEEDLE, 18 GAUGE 918201

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-07 for BARD BRACHYSTAR BRACHYTHERAPY APPLICATOR NEEDLE, 18 GAUGE 918201 manufactured by Bard Brachytherapy, Inc. -1424526.

Event Text Entries

[76948992] The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard. The device was not returned.
Patient Sequence No: 1, Text Type: N, H10


[76948993] It was reported that upon opening the package, the user found the needle was bent.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1018233-2017-02994
MDR Report Key6620459
Date Received2017-06-07
Date of Report2017-10-26
Date of Event2016-07-26
Date Mfgr Received2017-10-24
Device Manufacturer Date2016-01-20
Date Added to Maude2017-06-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactAMY GRAVLEY
Manufacturer Street8195 INDUSTRIAL BLVD
Manufacturer CityCOVINGTON GA 30014
Manufacturer CountryUS
Manufacturer Postal30014
Manufacturer Phone7707846100
Manufacturer G1BARD BRACHYTHERAPY, INC. -1424526
Manufacturer Street295 EAST LIES ROAD PARQUE INDUSTRIAL COLONIAL
Manufacturer CityCAROL STREAM IL 60188
Manufacturer CountryUS
Manufacturer Postal Code60188
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBARD BRACHYSTAR BRACHYTHERAPY APPLICATOR NEEDLE, 18 GAUGE
Generic NameBRACHY NEEDLE
Product CodeIWJ
Date Received2017-06-07
Catalog Number918201
Lot NumberREZL2332
Device Expiration Date2018-12-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBARD BRACHYTHERAPY, INC. -1424526
Manufacturer Address295 EAST LIES ROAD PARQUE INDUSTRIAL COLONIAL CAROL STREAM IL 60188 US 60188


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.