ANTI-GRAVITY TREADMILL M320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-08 for ANTI-GRAVITY TREADMILL M320 manufactured by Alterg, Inc..

Event Text Entries

[77534952] Alterg has determined that the failure was due to a faulty ic. The ic was identified as suspect previously and has been substituted with alternative in may 2016. A hazard analysis of the reported problem was performed tht indicated the potential for failure can ext to occur (b)(4) uses of the product. There have been no reported injuries due to this event. Alterg will address the faulty component in the field via repair notice to all distributors and offer free replacement parts.
Patient Sequence No: 1, Text Type: N, H10


[77534953] As reported by distributor: problem description: - new patient was in the m320, ready to start calibration - when pressing start, the treadmill did not calibrate but the treadmill went immediately to max speed. The belt did not stop when pulling the safety lanyard: emergency stop did not work! - patients shoe soles and shoe toes were seriously damaged by the abrasive belt. - the m320 could only be stopped by interrupting the power: this fast action by the therapist prevented the patient of being hurt. We did service the m300-0446 immediately: replacing the motor controller board 102647 did solve both problems: max. Speed and emergency stop.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007033433-2017-00001
MDR Report Key6621552
Date Received2017-06-08
Date of Report2017-01-06
Date of Event2017-04-04
Date Mfgr Received2017-04-18
Device Manufacturer Date2014-08-25
Date Added to Maude2017-06-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR MICHAEL PHILLIPS
Manufacturer Street48438 MILMONT DRIVE
Manufacturer CityFREMONT CA 94538
Manufacturer CountryUS
Manufacturer Postal94538
Manufacturer G1ALTERG, INC.
Manufacturer Street48438 MILMONT DRIVE
Manufacturer CityFREMONT CA 94538
Manufacturer CountryUS
Manufacturer Postal Code94538
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameANTI-GRAVITY TREADMILL
Generic NameMODEL M320
Product CodeBXB
Date Received2017-06-08
Model NumberM320
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALTERG, INC.
Manufacturer Address48438 MILMONT DRIVE FREMONT CA 94538 US 94538


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-06-08

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.