CONNECTOR U/ADAPIT STR 1 ARM NOBASE 50/C 004204-A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-08 for CONNECTOR U/ADAPIT STR 1 ARM NOBASE 50/C 004204-A manufactured by Vyaire Medical, Inc.

Event Text Entries

[77212893] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[77212894] The customer reported that? The oxygen port on this connector seems to have? Melted? According to the facility, causing the oxygen concentrator unable to maintain liter flow?. The patient did not change the adaptor in over one year. Once replaced with new adaptor, liter flow improved. "the patient felt like she could not breathe and thought something was wrong with the concentrator when she noticed the ball to increase liter flow would not rise at all staying below 1 liter?. The customer reported that the patient was not told to change the connector every 30 days. The customer reported that did not know this at the time, and now have seen this written on the package. The patient? Did not have any respiratory comprise at all".
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8030673-2017-00341
MDR Report Key6624645
Date Received2017-06-08
Date of Report2017-08-04
Date of Event2017-05-10
Date Mfgr Received2017-07-12
Date Added to Maude2017-06-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMINDY FABER
Manufacturer Street26125 N. RIVERWOODS BLVD
Manufacturer CityMETTAWA IL 60045
Manufacturer CountryUS
Manufacturer Postal60045
Manufacturer G1VYAIRE MEDICAL, INC
Manufacturer StreetCERRADA V NO.85 PARQUE INDUSTRIAL
Manufacturer CityMEXICALI BAJA CALIFORNIA NORTE
Manufacturer CountryMX
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCONNECTOR U/ADAPIT STR 1 ARM NOBASE 50/C
Generic NameCONNECTOR, AIRWAY (EXTENSION)
Product CodeBZA
Date Received2017-06-08
Catalog Number004204-A
Lot Number0000991207
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL, INC
Manufacturer AddressCERRADA V?A DE LA PRODUCCI?N NO. 85 PARQUE INDUSTRIAL MEXICALI BAJA CALIFORNIA NORTE MX


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-06-08