CS750 CLEAN STATION

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-09 for CS750 CLEAN STATION manufactured by Belimed Ag.

Event Text Entries

[77424174] Analysis of the data log revealed that the hospital-designated professional cssd operators repeated the following sequence of actions a total of 123 times over a period of three weeks and four days prior to the incident: on the occasions when the washer doors did not close properly due to slightly misaligned door guides, activating the system warning signal and error message, the operator selected the command to override the warnings; the door closed, either on its own or by application of manual force, and the washing process was conducted as specified. On additional occasions, the doors closed correctly and no error message nor warning signal was displayed by the device. On (b)(6) 2017, the door again became stuck. The operator on duty reached in to forcefully close the door by pulling it. It slammed shut and her hand was injured. The device is provided with a clear warning label with regard to the danger of suffering hand injuries by crushing if a hand is introduced between the moving parts of the door, i. E. , when the door is activated to be closed or when opening. This warning is also included in the operating instructions and the danger is discussed during distributor-conducted training of the hospital employees. These warnings were ignored by the operator who suffered the injury. In addition, both during training and in the operating instructions it is clearly stated that whenever a warning signal and a warning message are displayed on the machine and the control panel, the device may not be used and the supervisor must be informed immediately in order to initiate the troubleshooting procedure and have the failure remedied by qualified professional service personnel, if required. In accordance with the information recorded on the data log, these instructions were ignored a total of 123 times. The problem directly leading to the injury is therefore user error / misuse. The door guides were realigned on may 24, 2017, and the device is functioning properly.
Patient Sequence No: 1, Text Type: N, H10


[77424175] The operator of the washer attempted to close the washer door by means of the appropriate control. When it did not close she reached in to forcefully pull it shut, breaking a bone in her hand. She was taken to er and her hand was placed in a cast.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3001061128-2017-00001
MDR Report Key6629116
Date Received2017-06-09
Date of Report2017-06-09
Date of Event2017-05-10
Date Mfgr Received2017-05-12
Device Manufacturer Date2016-02-18
Date Added to Maude2017-06-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEFAN ROEMER
Manufacturer StreetGRIENBACHSTRASSE 11
Manufacturer CityZUG, ZUG 6300
Manufacturer CountrySZ
Manufacturer Postal6300
Manufacturer Phone414497888
Manufacturer G1BELIMED AG
Manufacturer StreetGRIENBACHSTRASSE 11
Manufacturer CityZUG, ZUG 6300
Manufacturer CountrySZ
Manufacturer Postal Code6300
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCS750 CLEAN STATION
Generic NameWASHER OF BODY WASTE RECEPTACLES
Product CodeFLH
Date Received2017-06-09
Model NumberCS750
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBELIMED AG
Manufacturer AddressGRIENBACHSTRASSE 11 ZUG, ZUG 6300 SZ 6300


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.