MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-09 for 75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CML-75LN manufactured by Charter Medical, Ltd..
[77617832]
The sample was not available for investigation. The root cause of the film crack could not be conclusively determined and the exact point in the process where the film crack occurred is unknown. There are other potential causes that could result in a film crack or tear: - film damage incurred during handling of a frozen bag. - a label attached to the bag surface could interfere with the flexibility of the film during cryogenic freezing and thawing processes. - overfilling the bag. An overfilled bag could create resistance during insertion and removal of the bag from the metal cassette. - excessive air left inside the bag after filling, resulting in rapid expansion during thawing. - moisture on the exterior surface of the bag or cassette interior when the bag is inserted into the metal cassette. Moisture may cause the bag film to freeze to the interior cassette surface, thus potentially resulting in damage when the frozen bag is removed from the cassette. - although there was no evidence, there is always the possibility that an undetected material issue could contribute to a tear of this nature. The product instruction for use sheet is provided with each cml-75ln device. Review of the ifu version provided with lot 142707 has the following precautions: - "do not write on bag surface, back side of label or adhere labels to bag surface. " - "after freezing, do not handle excessively. Port tubes and film are fragile in the frozen state and breakage may occur. Handle with care. " - "do not overfill. " - "remove as much air as possible from the container. " - "ensure bag exterior and protective freezing cassettes are dry prior to initiating freezing protocol. Moisture on the exterior of the bag or on the cassette could cause adherence of the bag to the cassette resulting in difficulty of bag removal. "
Patient Sequence No: 1, Text Type: N, H10
[77617833]
The bag break was discovered during the thawing process while the cell material was being processed. There was no contact with a patient when the leak occurred. There was no injury, death or further medical intervention required as a result of this occurrence. There was no medical staff exposure as a result of this occurrence.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1066733-2017-00006 |
MDR Report Key | 6631075 |
Date Received | 2017-06-09 |
Date of Report | 2017-06-08 |
Date of Event | 2017-04-18 |
Date Mfgr Received | 2017-05-09 |
Device Manufacturer Date | 2015-08-20 |
Date Added to Maude | 2017-06-09 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. JESSICA HUGHES |
Manufacturer Street | 3948-A WESTPOINT BLVD |
Manufacturer City | WINSTON SALEM NC 27103 |
Manufacturer Country | US |
Manufacturer Postal | 27103 |
Manufacturer Phone | 3367686447 |
Manufacturer G1 | CHARTER MEDICAL, LTD. |
Manufacturer Street | 3948-A WESTPOINT BLVD. |
Manufacturer City | WINSTON SALEM NC 27103 |
Manufacturer Country | US |
Manufacturer Postal Code | 27103 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | 75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER |
Generic Name | 75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER |
Product Code | LPZ |
Date Received | 2017-06-09 |
Model Number | CML-75LN |
Catalog Number | CML-75LN |
Lot Number | 142707 |
Device Expiration Date | 2018-08-01 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | CHARTER MEDICAL, LTD. |
Manufacturer Address | 3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-06-09 |