[77936414] The sample was not available for investigation. The root cause of the film crack could not be conclusively determined and the exact point in the process where the film crack occurred is unknown. There are other causes that could potentially result in a film crack or tear: film damage incurred during handling of a frozen bag. A label attached to the bag surface could interfere with the flexibility of the film during cryogenic freezing and thawing processes. Overfilling the bag. An overfilled bag could create resistance during insertion and removal of the bag from the metal cassette. Excessive air left inside the bag after filling, resulting in rapid expansion during thawing. Moisture on the exterior surface of the bag or cassette interior when the bag is inserted into the metal cassette. Moisture may cause the bag film to freeze to the interior cassette surface, thus potentially resulting in damage when the frozen bag is removed from the cassette. Although there was no evidence, there is always the possibility that an undetected material issue could contribute to a tear of this nature. The product instruction for use sheet is provided with each cml-75ln device. Review of the ifu version provided with lot 142707 has the following precautions: "do not write on bag surface, back side of label or adhere labels to bag surface. " "after freezing, do not handle excessively. Port tubes and film are fragile in the frozen state and breakage may occur. Handle with care. " "do not overfill. " "remove as much air as possible from the container. " "ensure bag exterior and protective freezing cassettes are dry prior to initiating freezing protocol. Moisture on the exterior of the bag or on the cassette could cause adherence of the bag to the cassette resulting in difficulty of bag removal. " sample not returned to charter medical.
Patient Sequence No: 1, Text Type: N, H10


[77936417] The bag break was discovered during the thawing process while the cell material was being processed. There was no contact with a patient when the leak occurred. There was no injury, death or further medical intervention required as a result of this occurrence. There was no medical staff exposure as a result of this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2017-00007
MDR Report Key6631080
Date Received2017-06-09
Date of Report2017-06-08
Date of Event2017-04-20
Date Mfgr Received2017-05-09
Device Manufacturer Date2015-08-20
Date Added to Maude2017-06-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA HUGHES
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Generic Name75-ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2017-06-09
Model NumberCML-75LN
Catalog NumberCML-75LN
Lot Number142707
Device Expiration Date2018-08-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.