MONOMAX VIOLET 1(4)150CM HR48 LOOP(M) B0041119

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-06-09 for MONOMAX VIOLET 1(4)150CM HR48 LOOP(M) B0041119 manufactured by B.braun Surgical Sa.

Event Text Entries

[77457584] (b)(4). Manufacturing site evaluation: samples received: 1 open pouch. Analysis and results: there is one previous complaint of the same code-batch but regarding other issue. Manufactured and distributed in the market (b)(4) units of this code batch. There are no units in stock. The sample received has the first pack sealed to second pack in the 4th sealing, as can be seen in enclosed picture. This defect took place in the welding machine and this unit was not sorted out by the personnel involved in this process. Final conclusion: taking into account that the sample received does not fulfill the oem specifications, it is concluded that the complaint is justified. Corrective/preventive actions: this complaint will be included in the analysis of trending, and corrective action will be open if applies.
Patient Sequence No: 1, Text Type: N, H10


[77457585] County of complaint: (b)(6). It is reported the monomax packaging is welded to the secondary packaging. When trying to remove the monomax packaging from the secondary packaging, it will risk sterilization..
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003639970-2017-00294
MDR Report Key6632054
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-06-09
Date of Report2017-06-09
Date of Event2017-05-17
Date Facility Aware2017-05-26
Date Mfgr Received2017-05-17
Date Added to Maude2017-06-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1B.BRAUN SURGICAL SA
Manufacturer Street121 CARRETERA DE TERRASSA
Manufacturer CityRUBI, 08191
Manufacturer CountrySP
Manufacturer Postal Code08191
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMONOMAX VIOLET 1(4)150CM HR48 LOOP(M)
Generic NameSUTURES
Product CodeNWJ
Date Received2017-06-09
Returned To Mfg2017-05-23
Model NumberB0041119
Catalog NumberB0041119
Lot Number116373V003
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerB.BRAUN SURGICAL SA
Manufacturer Address121 CARRETERA DE TERRASSA RUBI, 08191 SP 08191


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-06-09

© 2024 FDA.report