SET 60DROP STAT2 SPLIT SEP S2V-60

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-13 for SET 60DROP STAT2 SPLIT SEP S2V-60 manufactured by Conmed Corporation.

Event Text Entries

[77616861] The user facility, (b)(6) hospital, reported the device was not allowing fluid/drugs to drip properly. The device was returned, flow rate was tested during evaluation and the complaint was not confirmed. The device was able to regulate flow consistently, turn off and run full open properly. Variation in drip rate is expected with this product. The instruction for use cautions the user that dial numbers are approximate. Medical professionals should confirm drip rate by using timer or watch with a second hand. The product specification states there must be a continuous increase in flow rate when the dial is turned from off to full. This device met specification. The complaint product was re-evaluated as part of a root cause investigation on 08 jun 2017. Engineers disassembled the item to evaluate the internal drum. The drum channel was incorrect and is being addressed with a supplier corrective action request. Based on these findings, this complaint is considered confirmed and supplier related.
Patient Sequence No: 1, Text Type: N, H10


[77616862] The user facility reported the device was not allowing fluids/drugs to drip properly. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007305485-2017-00137
MDR Report Key6636866
Date Received2017-06-13
Date of Report2017-06-13
Date Mfgr Received2017-06-08
Device Manufacturer Date2015-07-14
Date Added to Maude2017-06-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSEY SHEPPARD
Manufacturer Street525 FRENCH ROAD
Manufacturer CityUTICA NY 135025994
Manufacturer CountryUS
Manufacturer Postal135025994
Manufacturer G1CONMED CORPORATION
Manufacturer Street525 FRENCH ROAD
Manufacturer CityUTICA NY 135025994
Manufacturer CountryUS
Manufacturer Postal Code135025994
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSET 60DROP STAT2 SPLIT SEP
Generic NameSET, ADMINISTRATION, INTRAVASCULAR
Product CodeLDR
Date Received2017-06-13
Returned To Mfg2015-11-20
Catalog NumberS2V-60
Lot Number201507144
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCONMED CORPORATION
Manufacturer Address525 FRENCH ROAD UTICA NY 135025994 US 135025994


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-13

© 2024 FDA.report