RECOVERY CONE REMOVAL SYSTEM RC15

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2017-06-13 for RECOVERY CONE REMOVAL SYSTEM RC15 manufactured by Bard Peripheral Vascular, Inc..

Event Text Entries

[77516936] (b)(4). Manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing. This lot met all release criteria. There was nothing found to indicate there was a manufacturing related cause for this event. This is the only event reported to date for this lot number and failure mode. Visual/microscopic inspection: the retrieval sheath was returned. The detached marker band was not returned with the retrieval device. The distal end of the retrieval sheath was examined under microscopic magnification. The distal marker band impression was identified on the appropriate location on the sheaths' surface. This indicates that the marker band was swaged on the retrieval sheath during manufacturing. A kink was noted in the sheath 33. 9cm from the proximal end of the hub. This kink most likely is due to the configuration in which the sheath was sent back for evaluation. As the user did not complain of the kink this will be an incidental finding. The distal tip of the sheath was also noted to be buckled. No other anomalies were identified on the sheath. Medical records review: no medical records have been made available to the manufacturer. Image/photo review: no medical images have been made available to the manufacturer. Conclusion: the device was returned for evaluation. One retrieval sheath was received without a distal marker band. The retrieval sheath clearly showed where the marker band was originally swaged onto the device. The detached marker band was not returned for evaluation. Therefore, the investigation is confirmed for the detached marker band. It should be noted that per the reported event, the recovery cone was used to retrieve a denali filter. Per the denali ifu (instructions for use), the denali filter should be removed using an intravascular loop snare only. The recovery cone is only indicated for use to percutaneously remove the g2 x filter, g2 express filter, g2 filter and the recovery filter from the vena cava, or facilitate the retrieval of foreign objects from the peripheral vascular system. The recovery cone is not indicated for use to remove a denali filter. Therefore, it is likely that user related issues (use of a recovery cone to remove a denali filter) contributed to the alleged marker band detachment. Labeling review: the current ifu (instructions for use) states: general information: the recovery cone removal system is intended to percutaneously remove the g2 x filter, g2 express filter, g2 filter, recovery filter or a foreign body as indicated. Indications for use: the recovery cone removal system is intended for use to percutaneously remove the g2 x filter, g2 express filter, g2 filter and the recovery filter from the vena cava, or facilitate the retrieval of foreign objects from the peripheral vascular system. Warnings: do not use excessive force when manipulating the cone. Excessive force may damage the catheter or other parts of the recovery cone removal system. Equipment required: - 12 french dilator directions for use - g2 x filter, g2 express filter, g2 filter or recovery filter removal - insert the guidewire and gently advance it to the location of the g2 x filter, g2 express filter, g2 filter or recovery filter for removal. - pre-dilate the accessed vessel with a 12 french dilator. - advance the 10 french introducer catheter together with its tapered dilator over the guidewire and into the vein, such that the tip of the sheath is approximately 3cm cephalad to the filter tip. Note: the introducer catheter has a radiopaque marker at the distal end of the catheter sheath to assist in visualization. - perform a standard inferior venacavogram (typically 30 ml of contrast medium at 15 ml/s). Check for thrombus within the filter. If there is significant thrombus within the filter, do not remove the g2 x filter, g2 express filter, g2 filter or recovery filter. Capture and removal of the g2 x filter, g2 express filter, g2 filter or the recovery filter: - the capture of the g2 x filter, g2 express filter, g2 filter or recovery filter is illustrated in the ifu (instructions for use). The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
Patient Sequence No: 1, Text Type: N, H10


[77516937] It was reported approximately six months post filter deployment, during a filter retrieval procedure, a vena cava filter removal system was used to successfully capture and remove the filter; however, the radiopaque marker band detached from the sheath and embolized to the pulmonary artery. The marker band was unable to be retrieved. There was no reported patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020394-2017-00616
MDR Report Key6639215
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2017-06-13
Date of Report2017-06-13
Date of Event2017-05-18
Date Mfgr Received2017-05-18
Device Manufacturer Date2017-03-17
Date Added to Maude2017-06-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJUDITH LUDWIG
Manufacturer Street1625 W 3RD ST.
Manufacturer CityTEMPE AZ 85281
Manufacturer CountryUS
Manufacturer Postal85281
Manufacturer Phone4803032689
Manufacturer G1C.R. BARD, INC. (GFO)
Manufacturer Street289 BAY ROAD
Manufacturer CityQUEENSBURY NY 12804
Manufacturer CountryUS
Manufacturer Postal Code12804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRECOVERY CONE REMOVAL SYSTEM
Generic NameVENA CAVA FILTER REMOVAL SYSTEM
Product CodeMMX
Date Received2017-06-13
Returned To Mfg2017-06-02
Catalog NumberRC15
Lot NumberGFBN3372
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBARD PERIPHERAL VASCULAR, INC.
Manufacturer Address1625 W 3RD ST. TEMPE AZ 85281 US 85281


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.