MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-06-15 for ADVIA 1800 manufactured by Siemens Healthcare Diagnostics Inc..
[77625723]
The customer contacted a siemens customer care center (ccc) and stated that a discordant lipase result was obtained on a patient sample, when the sample was run from the rack handler. The same sample was processed in the sample carousel and the result obtained matched the clinical picture of the patient. The cause of the discordant, falsely elevated lipase result on one patient sample is unknown. Siemens is investigating the issue.
Patient Sequence No: 1, Text Type: N, H10
[77625724]
A discordant, falsely elevated lipase result was obtained on one patient sample on an advia 1800 instrument. The initial result was not reported to the physician(s). The sample was repeated on the same instrument, resulting lower and matching the clinical picture of the patient. The repeat result was reported to the physician(s). There are no reports of patient intervention or adverse health consequence due to the discordant, falsely elevated lipase result.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2432235-2017-00365 |
MDR Report Key | 6642725 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2017-06-15 |
Date of Report | 2017-07-27 |
Date of Event | 2017-05-19 |
Date Mfgr Received | 2017-05-31 |
Device Manufacturer Date | 2013-11-28 |
Date Added to Maude | 2017-06-15 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | KARL AEBIG |
Manufacturer Street | 511 BENEDICT AVENUE |
Manufacturer City | TARRYTOWN NY 10591 |
Manufacturer Country | US |
Manufacturer Postal | 10591 |
Manufacturer Phone | 9145243102 |
Manufacturer G1 | JEOL LTD. (REGISTRATION # 3003637681) |
Manufacturer Street | 3-1-2 MUSASHINO AKISHIMA |
Manufacturer City | TOKYO, 196-8558 |
Manufacturer Country | JA |
Manufacturer Postal Code | 196-8558 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ADVIA 1800 |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | CHI |
Date Received | 2017-06-15 |
Model Number | ADVIA 1800 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
Manufacturer Address | 511 BENEDICT AVE TARRYTOWN NY 10591 US 10591 |
Brand Name | ADVIA 1800 |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2017-06-15 |
Model Number | ADVIA 1800 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
Manufacturer Address | 511 BENEDICT AVE TARRYTOWN NY 10591 US 10591 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-06-15 |