BIPRO SYSTEMS SDS CB200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-14 for BIPRO SYSTEMS SDS CB200 manufactured by Medical Solution.

Event Text Entries

[77647202] Hemodialysis pts c/o nausea, vomit, hypotension in between change over of 1st and 2nd shift 10:00 am - 12:00pm. Staff identified bicarb tank. Malfunction in which bicarb in day tank was diluted, bicarb conductivity out of range, indicating dilution.
Patient Sequence No: 1, Text Type: D, B5


[77647203] Hemodialysis pts c/o nausea, vomit, hypotension in between change over of 1st and 2nd shift 10:00 am - 12:00pm. Staff identified bicarb tank. Malfunction in which bicarb in day tank was diluted, bicarb conductivity out of range, indicating dilution.
Patient Sequence No: 2, Text Type: D, B5


[77647204] Hemodialysis pts c/o nausea, vomit, hypotension in between change over of 1st and 2nd shift 10:00 am - 12:00pm. Staff identified bicarb tank. Malfunction in which bicarb in day tank was diluted, bicarb conductivity out of range, indicating dilution.
Patient Sequence No: 3, Text Type: D, B5


[77647205] Hemodialysis pts c/o nausea, vomit, hypotension in between change over of 1st and 2nd shift 10:00 am - 12:00pm. Staff identified bicarb tank. Malfunction in which bicarb in day tank was diluted, bicarb conductivity out of range, indicating dilution.
Patient Sequence No: 4, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5070418
MDR Report Key6643186
Date Received2017-06-14
Date of Report2017-06-14
Date of Event2017-06-07
Date Added to Maude2017-06-15
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBIPRO SYSTEMS SDS CB200
Generic NameMIXING AND DISTRIBUTION SYSTEM OF BICARBONATE FOR DIALYSIS
Product CodeFIN
Date Received2017-06-14
Model NumberSDS CB200
Device AvailabilityY
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerMEDICAL SOLUTION
Manufacturer Address5646 MERRIAM DRIVE MERRIAM KS 66203 US 66203


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-14
20 2017-06-14
30 2017-06-14
40 2017-06-14

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.