MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-14 for BIPRO SYSTEMS SDS CB200 manufactured by Medical Solution.
[77647202]
Hemodialysis pts c/o nausea, vomit, hypotension in between change over of 1st and 2nd shift 10:00 am - 12:00pm. Staff identified bicarb tank. Malfunction in which bicarb in day tank was diluted, bicarb conductivity out of range, indicating dilution.
Patient Sequence No: 1, Text Type: D, B5
[77647203]
Hemodialysis pts c/o nausea, vomit, hypotension in between change over of 1st and 2nd shift 10:00 am - 12:00pm. Staff identified bicarb tank. Malfunction in which bicarb in day tank was diluted, bicarb conductivity out of range, indicating dilution.
Patient Sequence No: 2, Text Type: D, B5
[77647204]
Hemodialysis pts c/o nausea, vomit, hypotension in between change over of 1st and 2nd shift 10:00 am - 12:00pm. Staff identified bicarb tank. Malfunction in which bicarb in day tank was diluted, bicarb conductivity out of range, indicating dilution.
Patient Sequence No: 3, Text Type: D, B5
[77647205]
Hemodialysis pts c/o nausea, vomit, hypotension in between change over of 1st and 2nd shift 10:00 am - 12:00pm. Staff identified bicarb tank. Malfunction in which bicarb in day tank was diluted, bicarb conductivity out of range, indicating dilution.
Patient Sequence No: 4, Text Type: D, B5
Report Number | MW5070418 |
MDR Report Key | 6643186 |
Date Received | 2017-06-14 |
Date of Report | 2017-06-14 |
Date of Event | 2017-06-07 |
Date Added to Maude | 2017-06-15 |
Event Key | 0 |
Report Source Code | Voluntary report |
Manufacturer Link | N |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | BIPRO SYSTEMS SDS CB200 |
Generic Name | MIXING AND DISTRIBUTION SYSTEM OF BICARBONATE FOR DIALYSIS |
Product Code | FIN |
Date Received | 2017-06-14 |
Model Number | SDS CB200 |
Device Availability | Y |
Device Eval'ed by Mfgr | I |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDICAL SOLUTION |
Manufacturer Address | 5646 MERRIAM DRIVE MERRIAM KS 66203 US 66203 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-06-14 | |
2 | 0 | 2017-06-14 | |
3 | 0 | 2017-06-14 | |
4 | 0 | 2017-06-14 |