ARCHITECT CEA 07K68-27

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-15 for ARCHITECT CEA 07K68-27 manufactured by Abbott Ireland.

Event Text Entries

[77735326] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[77735327] The customer reported they obtained a false depressed result on the architect cea assay with reagent lot 72324fn00. Sample id (b)(6) generated an initial result of 3. 35 ng/ml and when repeated in duplicate (sample id (b)(6)) generated 149. 00 and 142. 16 ng/ml.
Patient Sequence No: 1, Text Type: D, B5


[96236997] Update: patient and device codes were provided.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3008344661-2017-00057
MDR Report Key6645040
Date Received2017-06-15
Date of Report2017-07-13
Date of Event2017-06-07
Date Mfgr Received2017-07-07
Device Manufacturer Date2017-02-10
Date Added to Maude2017-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT IRELAND
Manufacturer StreetDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK
Manufacturer CitySLIGO NA
Manufacturer CountryEI
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT CEA
Generic NameCARCINOEMBRYONIC ANTIGEN
Product CodeDHX
Date Received2017-06-15
Catalog Number07K68-27
Lot Number72324FN00
Device Expiration Date2018-07-23
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT IRELAND
Manufacturer AddressDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK SLIGO NA EI NA


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.