STERRAD? SEALSURE CHEMICAL INDICATOR TAPE 14202_90

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-06-15 for STERRAD? SEALSURE CHEMICAL INDICATOR TAPE 14202_90 manufactured by Advanced Sterilization Products.

Event Text Entries

[77720330] A customer reported using expired sterrad? Chemical indicator tape to process a sterrad? Cycle. The affected load was released and used on patients. There was no report of infection, injury or harm to patient(s) associated with this issue. Although there is no report of patient injury or harm and no prior incidents have resulted in serious injury, advanced sterilization products (asp) has determined in this situation sterility cannot be assured.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2084725-2017-00501
MDR Report Key6645766
Report SourceHEALTH PROFESSIONAL
Date Received2017-06-15
Date of Report2017-05-26
Date of Event2017-05-26
Date Mfgr Received2017-07-23
Date Added to Maude2017-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JOAQUIN KURZ
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone6506874924
Manufacturer G1ADVANCED STERILIZATION PRODUCTS
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTERRAD? SEALSURE CHEMICAL INDICATOR TAPE
Generic NameINDICATOR, CHEMICAL
Product CodeJOJ
Date Received2017-06-15
Model NumberNA
Catalog Number14202_90
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerADVANCED STERILIZATION PRODUCTS
Manufacturer Address33 TECHNOLOGY IRVINE 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.