ARTICULATING ARM VP8525

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2017-06-16 for ARTICULATING ARM VP8525 manufactured by Conmed Corportation.

Event Text Entries

[77773670] At this time conmed is awaiting receipt of the device for an evaluation. Upon the completion of the product evaluation and complaint investigation a supplemental and final report will be filed.
Patient Sequence No: 1, Text Type: N, H10


[77773671] The sales representative reported that the user facility contacted him on (b)(6) 2017 to report the articulating arm fell off today and landed on a nurse in the operating room. The nurse is on medical leave. Follow-up with the user facility was made on june 6, 2017. According to the nurse manager, as the nurse was removing the cover on a 40" monitor being held up by the articulating arm bracket the monitor fell off the bracket. The customer has advised that the bracket holes of the articulating arm were upside down on the plate that the screws are installed through. The nurse twisted to try to catch the monitor and wrenched her back. The nurse spent time at home with limited activity and is now fully recovered.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017294-2017-00060
MDR Report Key6647465
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2017-06-16
Date of Report2017-08-02
Date of Event2017-05-30
Date Mfgr Received2017-08-01
Device Manufacturer Date2015-11-30
Date Added to Maude2017-06-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSEY SHEPPARD
Manufacturer Street525 FRENCH ROAD
Manufacturer CityUTICA NY 135025994
Manufacturer CountryUS
Manufacturer Postal135025994
Manufacturer Phone7273995209
Manufacturer G1CONMED CORPORATION
Manufacturer Street525 FRENCH ROAD
Manufacturer CityUTICA NY 135025994
Manufacturer CountryUS
Manufacturer Postal Code135025994
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARTICULATING ARM
Generic NameCAMERA, SURGICAL AND ACCESSORIES
Product CodeKQM
Date Received2017-06-16
Returned To Mfg2017-07-25
Catalog NumberVP8525
Lot Number150716B
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCONMED CORPORTATION
Manufacturer Address11311 CONCEPT BOULEVARD LARGO FL 33773 US 33773


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-06-16

© 2024 FDA.report