STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903100 390-310

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-06-16 for STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903100 390-310 manufactured by Boston Scientific - Marlborough.

Event Text Entries

[77834268] (b)(4). A visual analysis of the returned device found the device was returned in one piece. The core wire pulled out when trying to close the device. Furthermore, the core wire was broken into two pieces. The blue sheath and white heat shrink were found accordion. Moreover, the white shrink was torn. In addition, the blue/ green shrink was pulled away from distal stop. Evaluation concluded that the broken core wire most likely caused by the lithotripter during the procedure. Therefore, the most probable root cause for this complaint event is caused by other device. A review of the device history record (dhr) confirms that the accepted device met all manufacturing specifications.
Patient Sequence No: 1, Text Type: N, H10


[77834269] It was reported to boston scientific corporation that a stone cone nitinol urological retrieval coil was used in the ureter during a ureterolithotomy procedure on (b)(6) 2017. According to the complainant, during the procedure, the handle end of the blue sheath accordioned when attempting to close the device. The coil failed to close and there was difficulty removing the device from the ureter. Device was disassembled and was removed from the patient. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be stable. This event has been deemed a reportable event based on the investigation results; core wire broken.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2017-01763
MDR Report Key6649024
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-06-16
Date of Report2017-05-23
Date of Event2017-03-28
Date Mfgr Received2017-05-23
Device Manufacturer Date2015-05-08
Date Added to Maude2017-06-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1ACCELLENT ENDOSCOPY JUAREZ
Manufacturer Street31C BUTTERFIELD TRAIL
Manufacturer CityEL PASO TX 79906
Manufacturer CountryUS
Manufacturer Postal Code79906
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTONE CONE NITINOL UROLOGICAL RETRIEVAL COIL
Generic NameDISLODGER, STONE, FLEXIBLE
Product CodeFGO
Date Received2017-06-16
Returned To Mfg2017-04-19
Model NumberM0063903100
Catalog Number390-310
Lot NumberD86221
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - MARLBOROUGH
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.