EXPRO ELITE SNARE 6600-015

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2017-06-19 for EXPRO ELITE SNARE 6600-015 manufactured by Vascular Solutions, Inc.

Event Text Entries

[77843658] This event occured in (b)(6). Per us: similar device identified as: brand name: expro elitre snare. Common-device name: snare. The 510(k) 071457. One expro elite snare was returned to vsi for evaluation. The loop coil had broken just distal to the intact joint coil. As no non-conformities were found in the manufacturing records and the most likely root cause is excessive force applied by the user, no corrective action is required.
Patient Sequence No: 1, Text Type: N, H10


[77843659] Physician tried to catch a line of port catheter with expro elite snare, by doing this the tip of the snare was lost in the body of the patient. The catheter was retrieved and tip of snare was left in patient. Patient recovered without symptoms. Percutaneous intervention was performed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134812-2017-00057
MDR Report Key6649676
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2017-06-19
Date of Report2016-05-03
Date of Event2016-05-02
Date Mfgr Received2016-05-03
Device Manufacturer Date2015-02-28
Date Added to Maude2017-06-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS MARY HAUFEK
Manufacturer Street6464 SYCAMORE COURT NORTH
Manufacturer CityMINNEAPOLIS MN 55369
Manufacturer CountryUS
Manufacturer Postal55369
Manufacturer Phone7636564230
Manufacturer G1VASCULAR SOLUTIONS, INC
Manufacturer Street6464 SYCAMORE COURT NORTH
Manufacturer CityMINNEAPOLIS MN 55369
Manufacturer CountryUS
Manufacturer Postal Code55369
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEXPRO ELITE SNARE
Generic NameSNARE
Product CodeMMX
Date Received2017-06-19
Returned To Mfg2016-09-22
Model Number6600-015
Lot Number581242
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVASCULAR SOLUTIONS, INC
Manufacturer Address6464 SYCAMORE COURT NORTH MINNEAPOLIS MN 55369 US 55369


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-06-19