STAT PROFILE PRIME CCS ANALYZER WITH SCANNER 2 53657

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-19 for STAT PROFILE PRIME CCS ANALYZER WITH SCANNER 2 53657 manufactured by Nova Biomedical Corporation.

Event Text Entries

[78029773] A review of the customer complaint information confirmed that the patient was not mistreated due to the identified software issue. Nova has also confirmed that this is the first complaint that we have received due to this reported software issue. Decision to initiate a field correction to immediately correct the software issue associated with the use of stat profile prime auto-cartridge quality control (a-qc) materials. The investigation is currently in process.
Patient Sequence No: 1, Text Type: N, H10


[78029774] Prime analyzer s/n (b)(4) failing to flag a patient critical ph level.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1219029-2017-00006
MDR Report Key6651879
Date Received2017-06-19
Date of Report2017-05-22
Date of Event2017-05-18
Date Mfgr Received2017-05-19
Device Manufacturer Date2015-11-24
Date Added to Maude2017-06-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. RACHEL KULA
Manufacturer Street200 PROSPECT STREET
Manufacturer CityWALTHAM MA 02453
Manufacturer CountryUS
Manufacturer Postal02453
Manufacturer Phone7816473700
Manufacturer G1NOVA BIOMEDICAL CORPORATION
Manufacturer Street200 PROSPECT ST
Manufacturer CityWALTHAM MA 024533465
Manufacturer CountryUS
Manufacturer Postal Code024533465
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSTAT PROFILE PRIME CCS ANALYZER WITH SCANNER 2
Generic NameANALYZER
Product CodeGKF
Date Received2017-06-19
Model Number53657
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNOVA BIOMEDICAL CORPORATION
Manufacturer Address200 PROSPECT ST WALTHAM MA 02454 US 02454


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-19