[78053032] (b)(4). Based on risk assessment conducted may 25, 2017, it was determined that there is a risk of injury if the malfunction were to recur. There have been no reported injuries. Reference field corrective action number: 3010611950/06/09/2017/001-r.
Patient Sequence No: 1, Text Type: N, H10


[78053033] Excessive static noise observed in elite probe.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010611950-2017-00014
MDR Report Key6656721
Report SourceDISTRIBUTOR
Date Received2017-06-21
Date of Report2017-03-23
Date of Event2017-03-23
Date Mfgr Received2017-03-23
Device Manufacturer Date2017-02-13
Date Added to Maude2017-06-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationHOSPITAL SERVICE TECHNICIAN
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JANESSA BOONE
Manufacturer Street3150 PLEASANT VIEW ROAD
Manufacturer CityMIDDLETON WI 53562
Manufacturer CountryUS
Manufacturer Postal53562
Manufacturer Phone6088298603
Manufacturer G1NATUS NEUROLOGY INCORPORATED
Manufacturer Street3150 PLEASANT VIEW ROAD
Manufacturer CityMIDDLETON WI 53562
Manufacturer CountryUS
Manufacturer Postal Code53562
Single Use3
Remedial ActionRC
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELITE PROBE
Generic NameELITE 3MHZ PROBE
Product CodeJAF
Date Received2017-06-21
Returned To Mfg2017-05-25
Model NumberX1L001
Catalog NumberN300
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS NEUROLOGY, INCORPORATED
Manufacturer Address3150 PLEASANT VIEW ROAD MIDDLETON WI 53562 US 53562


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-06-21

© 2026 FDA.report