S-L RETRACTOR TRANS MULTI-L 150MM RACK S-0103

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-21 for S-L RETRACTOR TRANS MULTI-L 150MM RACK S-0103 manufactured by Carefusion, Inc.

Event Text Entries

[78249855] (b)(4). Should any additional information be obtained for this issue a follow up emdr will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[78249856] The customer reported, during surgery the doctor was getting ready to use the device and realized that the product was not staying locked. Immediately stopped using and was able to use other one on hand. The key portion that helps it move in and out that makes it adjustable does not stay in place, making it fall in and out. The key that ratchets the retractor has a small screw directly in the end of it. That screw came out while opening the retractor on several cases. It went onto the field and had to be retrieved. No patient harm. No patient or user harm and no medical intervention required. Update 21jun2017 after multiple attempts, no additional information was received from the customer.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1923569-2017-00048
MDR Report Key6657422
Date Received2017-06-21
Date of Report2017-06-27
Date of Event2017-05-24
Date Mfgr Received2017-06-21
Date Added to Maude2017-06-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNA WEHRHEIM
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer G1CAREFUSION, INC
Manufacturer Street5 SUNNEN DR
Manufacturer CityST. LOUIS MO 63143
Manufacturer CountryUS
Manufacturer Postal Code63143
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameS-L RETRACTOR TRANS MULTI-L 150MM RACK
Generic NameRETRACTOR, SELF-RETAINING, FOR NEUROSURGERY
Product CodeGZT
Date Received2017-06-21
Model NumberS-0103
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address75 NORTH FAIRWAY DRIVE VERNON HILLS IL 60061 US 60061


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-21

© 2026 FDA.report