OAKWORKS MEDICAL CFPM 300 IMAGING TABLE 75209-T01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-06-22 for OAKWORKS MEDICAL CFPM 300 IMAGING TABLE 75209-T01 manufactured by Oakworks, Inc.,.

Event Text Entries

[78186237] Recurring issue.
Patient Sequence No: 1, Text Type: N, H10


[78186238] (b)(6) from performance imaging technologies called in and stated that they kick the foot pedal while moving around the table and in some instances, the foot pedal gets kicked under the base of the table. The foot pedal will get lodged under the base and the button engages. This has occurred with patients on the table. The last time this happened, they were getting ready to perform a lumbar injection when the foot pedal was kicked under the base. This engaged the lateral roll movement and the table started to roll. They were able to stop it in time and there was no harm to the patient. When asked if the patient was strapped down using the patient safety strap that comes with the table, the customer said no, they do not use the straps typically due to the location of the injections within the spinal column. (b)(6) was advised to unplug the foot pedal and to discontinue the use of it. They can use the hand control to operate the table.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2529571-2017-00005
MDR Report Key6660012
Report SourceHEALTH PROFESSIONAL
Date Received2017-06-22
Date of Report2017-06-06
Date of Event2015-03-31
Date Mfgr Received2015-03-31
Device Manufacturer Date2012-12-17
Date Added to Maude2017-06-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEPHEN MCKINLEY
Manufacturer Street923 EAST WELLSPRING RD
Manufacturer CityNEW FREEDOM PA 17349
Manufacturer CountryUS
Manufacturer Postal17349
Manufacturer G1OAKWORKS, INC.,
Manufacturer Street923 EAST WELLSPRING RD
Manufacturer CityNEW FREEDOM PA 17349
Manufacturer CountryUS
Manufacturer Postal Code17349
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOAKWORKS MEDICAL CFPM 300 IMAGING TABLE
Generic NameIMAGING TABLE
Product CodeIXR
Date Received2017-06-22
Model NumberCFPM 300
Catalog Number75209-T01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOAKWORKS, INC.,
Manufacturer Address923 EAST WELLSPRING RD NEW FREEDOM PA 17349 US 17349


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-22