OAKWORKS MEDICAL CFPM 400 IMAGING TABLE 75204-T01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2017-06-22 for OAKWORKS MEDICAL CFPM 400 IMAGING TABLE 75204-T01 manufactured by Oakworks, Inc.,.

Event Text Entries

[78192373]
Patient Sequence No: 1, Text Type: N, H10


[78192374] The foot pedal buttons are sticking due to damage caused by the column being lowered onto the foot pedal. The customer is using the hand control. There was no patient harm or adverse event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2529571-2017-00008
MDR Report Key6660022
Report SourceDISTRIBUTOR
Date Received2017-06-22
Date of Report2017-06-06
Date of Event2015-10-01
Date Mfgr Received2015-10-01
Device Manufacturer Date2013-08-16
Date Added to Maude2017-06-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEPHEN MCKINLEY
Manufacturer Street923 EAST WELLSPRING RD
Manufacturer CityNEW FREEDOM PA 17349
Manufacturer CountryUS
Manufacturer Postal17349
Manufacturer G1OAKWORKS, INC.,
Manufacturer Street923 EAST WELLSPRING RD
Manufacturer CityNEW FREEDOM PA 17349
Manufacturer CountryUS
Manufacturer Postal Code17349
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOAKWORKS MEDICAL CFPM 400 IMAGING TABLE
Generic NameIMAGING TABLE
Product CodeIXR
Date Received2017-06-22
Model NumberCFPM 400
Catalog Number75204-T01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOAKWORKS, INC.,
Manufacturer Address923 EAST WELLSPRING RD NEW FREEDOM PA 17349 US 17349


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-22