CHROMID? CPS ELITE AGAR 416172

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-06-22 for CHROMID? CPS ELITE AGAR 416172 manufactured by Biom?rieux Sa.

Event Text Entries

[78267796] A biom? Rieux internal investigation was perfromed. Innoculation of some additional species of citrobacter spp and hafnia alvei strains coming from atcc collection were performed, considering that some atypical strains such as citrobacter and h. Alvei could give pink coloration. These include: citrobacter freundii atcc 8090, citrobacter freundii atcc 8144101, hafnia alvei atcc 29926, hafnia alvei atcc 8208040. After inoculation, the sensitivity of identifying e. Coli according to the color and morphology of the colonies was excellent for the batch complaint after incubation for 18-24 hours , thereby, e. Coli strains produced the expected colony appearance, developing as spontaneous red to burgundy coloration, producing beta-glucuronidase (beta-gur) and/ or beta-galactosidase (beta-gur). Moreover, kesc strains, including both additional strains of citrobacter spp, produced the expected colony appearance developing as spontaneous blue to green coloration producing beta-glucosidase ( beta-glu). Both challenge strains for h. Alvei that were tested appeared as colorless colonies , not able produce acid from lactose or d-sorbitol and failure to elaborate the enzyme beta-d-glucuronidase. Based on these results, the problem was not confirmed during the investigation with the retention samples and the internal strain collection, nevertheless, it is common for most of h. Alvei strains to develop as colorless colonies; therefore, the different color among the customer strain and both h. Alvei internal strains (atcc 8208040 and atcc 29926) could be explained because this particular strain was isolated from a urinary sample and showed atypical enzymatic pathways expressing bet]-gur activity, confirming that problem is not batch or product related. Seems to be related with the particular clinical strain. Nevertheless, the specificity issue with h. Alvei is already known. It has been detected during the cps elite clinical study. So it is part of the specificity of the medium for some particular strains. This is the reason why it is included in limitations in the product insert (specificity of coloration for e. Coli of 97%).
Patient Sequence No: 1, Text Type: N, H10


[78267797] A customer from (b)(6) reported to biomerieux discrepant results when testing a patient urine sample with chromid? Cps elite agar (cpse) ref 416172, lot 1005585060. The customer tested the urine sample on cpse and observed a colony with color, characteristic of escherichia coli. The customer then performed susceptibility testing with vitek? 2 and found discrepant resistance. The sample was then tested with the vitek? 2 gn card and was identified as hafnia alvei at 98-99%. The customer reported no incorrect result was reported to the physician; however there was a delay greater than 24 hours due to the additional tests performed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00145
MDR Report Key6661146
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-06-22
Date of Report2017-06-22
Date Mfgr Received2017-05-23
Device Manufacturer Date2017-03-22
Date Added to Maude2017-06-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOM
Manufacturer Street5, RUE DES AQUEDUCS
Manufacturer CityCRAPONNE, FR 69290
Manufacturer CountryFR
Manufacturer Postal Code69290
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? CPS ELITE AGAR
Generic NameCHROMID? CPS ELITE AGAR
Product CodeJXA
Date Received2017-06-22
Catalog Number416172
Lot Number1005585060
ID Number03573026443900
Device Expiration Date2017-07-18
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX SA
Manufacturer Address5, RUE DES AQUEDUCS CRAPONNE, FR 69290 FR 69290


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.