STATIC AND DYNAMIC COMPLIANCE VMAX ENCORE 22D 777405-101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-06-22 for STATIC AND DYNAMIC COMPLIANCE VMAX ENCORE 22D 777405-101 manufactured by Carefusion, Inc.

Event Text Entries

[78355340] (b)(4). The customer reported the suspect component is available for analysis and an return good authorization (rga) has been issued. At this time, carefusion has not received the suspect component for evaluation. If additional information becomes available, it will be submitted in a follow up report.
Patient Sequence No: 1, Text Type: N, H10


[78355341] The customer reported while using the vmax encore, a smell of smoke was noticeable while warming up the unit. The issue occurred with no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


[113563480] The vyaire failure analysis lab received the suspected device and performed a failure investigation. The failure analysis technician confirmed burnt tracings on the analyzer control board (777348-101). The root cause is the operator error-assembly and is an isolated incident.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2021710-2017-06170
MDR Report Key6661466
Date Received2017-06-22
Date of Report2017-11-03
Date of Event2017-05-30
Date Mfgr Received2017-10-31
Device Manufacturer Date2014-06-01
Date Added to Maude2017-06-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKRISTIN GRAF
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer G1CAREFUSION, INC
Manufacturer Street1100 BIRD CENTER DR.
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSTATIC AND DYNAMIC COMPLIANCE
Generic NameCALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Product CodeBTY
Date Received2017-06-22
Model NumberVMAX ENCORE 22D
Catalog Number777405-101
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.