SYSMEX SA-10 CR690716

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-06-23 for SYSMEX SA-10 CR690716 manufactured by Sysmex Ra Co. Ltd..

Event Text Entries

[78371882] Investigation by sysmex corporation in (b)(4) determined lightning caused an excessive voltage that exceeded the product standard to be applied to the power supply unit. The surge absorber of the board was damaged. There was no abnormality found in the manufacturing of the product. The board was replaced and the instrument was operational. Any incident with excessive heat has the potential to cause user harm. The pcb is made of flame resistant materials and covered with a metal shield, but excessive heat from the event poses a risk of inhalation of smoke and/or harmful vapors from melted materials. The pcb is not accessible to the user, and is grounded, reducing the potential for harm due to shock.
Patient Sequence No: 1, Text Type: N, H10


[78371883] The user reported a power outage occurred during a thunderstorm. The analyzer power remained on through the use of an uninterrupted power supply (ups). When the user tried to perform sample analysis, the sampler was not functioning. A sysmex field service engineer (fse) went on site and found the surge absorber on the printed circuit board (pcb) power supply was charred due to excessive heat.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1000515253-2017-00024
MDR Report Key6665586
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-06-23
Date of Report2017-06-23
Date of Event2017-02-20
Date Mfgr Received2017-02-20
Device Manufacturer Date2012-10-12
Date Added to Maude2017-06-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. PETER SHEARSTONE
Manufacturer Street577 APTAKISIC RD
Manufacturer CityLINCOLNSHIRE IL 60069
Manufacturer CountryUS
Manufacturer Postal60069
Manufacturer Phone2245439514
Manufacturer G1SYSMEX RA CO. LTD.
Manufacturer Street1850-3 HIROOKANOMURA
Manufacturer CitySHIOJIRI, NAGANO 399-0702
Manufacturer CountryJA
Manufacturer Postal Code399-0702
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYSMEX SA-10
Generic NameSA-10 SAMPLER
Product CodeJRO
Date Received2017-06-23
Model NumberSA-10
Catalog NumberCR690716
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSYSMEX RA CO. LTD.
Manufacturer Address1850-3 HIROOKANOMURA SHIOJIRI, NAGANO 399-0702 JA 399-0702


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.