MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-06-27 for COBAS 6000 C (501) MODULE C501 04745914001 manufactured by Roche Diagnostics.
[78478947]
(b)(4). (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[78478948]
The customer received questionable results for two patient samples. Sample 1 initial total protein result was 52. 4 g/l. The doctor called and questioned the initial result. The repeat result was 71. 3 g/l. Sample 2 initial albumin result was 13. 4 g/l and the repeat result was 28. 9 g/l. The erroneous results were reported outside the laboratory. There was no allegation of an adverse event. The total protein reagent lot number was 246322. The albumin reagent lot number was 241675. The expiration dates were requested but were not provided.
Patient Sequence No: 1, Text Type: D, B5
[118478067]
The field service representative found the vacuum system blocked and was repaired. A blockage would have diluted the sample due to insufficient cell rinse performance. The issue was resolved by the service actions.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1823260-2017-01340 |
MDR Report Key | 6669369 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2017-06-27 |
Date of Report | 2017-07-09 |
Date of Event | 2017-06-13 |
Date Mfgr Received | 2017-06-13 |
Date Added to Maude | 2017-06-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | HITACHI HIGH TECH CORP. |
Manufacturer Street | 882 ICHIGE HITACHINAKA NA |
Manufacturer City | IBARAKI 312-8504 |
Manufacturer Country | JA |
Manufacturer Postal Code | 312-8504 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COBAS 6000 C (501) MODULE |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | JGQ |
Date Received | 2017-06-27 |
Model Number | C501 |
Catalog Number | 04745914001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Brand Name | COBAS 6000 C (501) MODULE |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2017-06-27 |
Model Number | C501 |
Catalog Number | 04745914001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-06-27 |