TEC 7

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-06-28 for TEC 7 manufactured by Datex-ohmeda, Inc..

Event Text Entries

[78609601] (b)(6). A ge healthcare service representative performed a checkout of the equipment. The vaporizer was found to fail leak testing. According to the hospital, the vaporizer has been sent to a third party service organization.
Patient Sequence No: 1, Text Type: N, H10


[78609602] Per user filed medwatch report, mw5070188: "crna smelled anesthesia gases and thought it was the pt's airway but could not find a leak. She became dizzy, and asked for another clinician to come into the room. The isoflurane vaporizer was leaking. Vaporizer turned off. Anesthesia tech removed the vaporizer and placed a new one on the machine. After several minutes, the second vaporizer was noted to be leaking. Second clinician took over the case, first clinician put on oxygen by nasal canula and called employee health, treated in the emergency room. Machine, vaporizers, and patient circuit removed and sent to biomed. Another machine was placed on patient. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2112667-2017-01195
MDR Report Key6674167
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-06-28
Date of Report2017-06-28
Date of Event2017-04-28
Date Mfgr Received2017-06-05
Device Manufacturer Date2015-12-16
Date Added to Maude2017-06-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJOSEPH SELIGA
Manufacturer Street3000 N GRANDVIEW BLVD.
Manufacturer CityWAUKESHA WI
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTEC 7
Generic NameVAPORIZER
Product CodeCAD
Date Received2017-06-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDATEX-OHMEDA, INC.
Manufacturer Address3030 OHMEDA DR, MADISON, WI US


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.