COBAS 6000 C (501) MODULE C501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-06-30 for COBAS 6000 C (501) MODULE C501 manufactured by Roche Diagnostics.

Event Text Entries

[78980755] This event occurred in (b)(6). Unique identifier (udi)#: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[78980756] The customer complained of erroneous results for 3 patients tested for igg-2 tina-quant igg gen. 2 (igg-2) on a cobas 6000 c (501) module. Based on the data provided, erroneous igm-2 tina-quant igm gen. 2 (igm-2) results were also identified for 1 of these patients. The samples are sent to the c501 module from the modular pre-analytic (mpa) system. The mpa makes 2 aliquots and one is tested on the customer's c501 module and the other is sent to the customer's sister site. The sister site performs electrophoresis on the samples and repeats the patient samples on their c8000 system. The customer runs the sample from a cup on the c501 module and the sister site runs the sample from a tube. The samples have been sent back from the sister site in the tubes and repeated on the customer's c501 module and the repeat tests give the correct result. Patient 1 initial igg-2 result on the c501 module was 6. 48 g/l. This result was reported outside of the laboratory. The repeat from the c8000 system used at the sister site was 15. 5 g/l. On (b)(6) 2017 the sample was repeated on the c501 module and the result was 16. 53 g/l. The result was later amended and reported outside of the laboratory as 15. 5 g/l. The last preventive maintenance was the week of (b)(6) 2017. There was a slight fluid leak identified at the time that was slightly diluting chemistries. Service at that time did not resolve the issue. On (b)(6) 2017 patient 2 initial igg-2 result on the c501 module was 44. 23 g/l and the initial igm-2 result was 0. 23 with a data flag. On (b)(6) 2017 the sample was tested on a different c501 module and the igg-2 result was 17. 39 g/l and the igm-2 result was 0. 30 g/l. The sample was repeated on the c501 module in question and the igg-2 result was 16. 33 g/l. It is not known if the erroneous results were reported outside of the laboratory. This information has been requested. On (b)(6) 2017 patient 3 (female with date of birth of (b)(6) 1944) initial igg-2 result on the c501 module was 3. 34 g/l. This result was reported outside of the laboratory. The repeat from the c8000 system used at the sister site was 8. 5 g/l. On (b)(6) 2017 the sample was repeated on the c501 module and the result was 7. 78 g/l. The sample was also repeated on a different c501 module at the customer site and the result was 8. 05 g/l. The result was later amended and reported outside of the laboratory as 8. 5 g/l. The samples looked ok upon visual inspection. The results that were reported were corrected in a timely fashion. No reagent lot numbers or expiration dates were provided. After the initial point of contact, the customer complained about a 4th patient sample with erroneous crpl3 c-reactive protein gen. 3 (crpl3), crea2 and ureal urea/bun results on the c501 module. On (b)(6) 2017 the initial crpl3 result was 1. 81 (unit of measure not provided), the initial crea2 result was 5. 8 (unit of measure not provided), and the initial ureal result was 1. 0 (unit of measure not provided). On (b)(6) 2017 the sample was repeated on the c501 module and the crpl3 result was 66. 40, the crea2 result was 206. 8 and the ureal result was 26. 3. It is not known if the erroneous results were reported outside of the laboratory. This information has been requested. The field service engineer (fse) visited the customer site and checked the rinse levels of the instrument which were all ok. The gear pump pressure was slightly low and this was adjusted. Both sample and reagent lines were decontaminated. Several other parts of the instrument were checked and were found to be within specification. A precision check was run on the igg-2 assay and the results were ok. Quality controls (qc) were run on the c 501 module and were ok.
Patient Sequence No: 1, Text Type: D, B5


[118808450] The customer clarified that the erroneous results for patients 2 and 4 were reported outside of the laboratory, however, the corrected results were provided once repeat testing was completed. No patients were adversely affected since no action was taken based on the erroneous results. The customer states the instrument is performing much better. They have been running samples in duplicate for 1 week and have had no more discrepant results.
Patient Sequence No: 1, Text Type: N, H10


[120549427] The fse returned to the customer site on 06/26/2017 due to a different issue. The photometer lens holder was removed and the seal had become very brittle and hard. The seal was replaced. Alignment was checked and performance testing was successful. A photometer check was also acceptable. The customer mentioned a concern with the rinse unit mechanism. The fse checked this and all was ok. The customer ran calibration and qc and the results were acceptable.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2017-01379
MDR Report Key6680638
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-06-30
Date of Report2017-08-23
Date of Event2017-06-01
Date Mfgr Received2017-06-09
Date Added to Maude2017-06-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 6000 C (501) MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeCFQ
Date Received2017-06-30
Model NumberC501
Catalog NumberASKU
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 6000 C (501) MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-06-30
Model NumberC501
Catalog NumberASKU
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-30

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.