ETEST? AMPICILLIN AM 256 WW S30 412253

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-06-30 for ETEST? AMPICILLIN AM 256 WW S30 412253 manufactured by Biomerieux, Sa.

Event Text Entries

[78984405] A customer from (b)(6) reported to biomerieux false susceptible ampicillin results in association with etest? Ampicillin am 256 ww s30. The customer reported that they compared results of patient enterococcus faecalis strains for etest? And disk diffusion methods. The customer stated that ten (10) strains of the forty-one (41) tested were susceptible with etest? Ampicillin and resistant with disk diffusion. The internal qc (e. Faecalis atcc 29212) for disk diffusion was in the range according to eucast but outside for etest? (0. 38mg/l for a target 0. 5 - 2. 0). The customer reported that patient results were not affected as disk diffusion was used as first line and there was no report of patient treatment being affected. There was no delay for reporting patient results. There is no indication or report from the hospital or treating physician to biom? Rieux that the discrepant result led to any adverse event related to the patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00153
MDR Report Key6682549
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-06-30
Date of Report2018-01-11
Date Mfgr Received2017-12-19
Device Manufacturer Date2016-10-11
Date Added to Maude2017-06-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOMERIEUX, SA
Manufacturer Street3 ROUTE DE PORT MICHAUD
Manufacturer CityLA BALME LES GROTTES ISERE, 38390
Manufacturer CountryFR
Manufacturer Postal Code38390
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETEST? AMPICILLIN AM 256 WW S30
Generic NameETEST? AMPICILLIN
Product CodeJWY
Date Received2017-06-30
Catalog Number412253
Lot Number1005224490
ID Number03573026376710
Device Expiration Date2019-10-11
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME LES GROTTES ISERE, 38390 FR 38390


Patients

Patient NumberTreatmentOutcomeDate
10 2017-06-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.