CELLEBRITY? M00516011 1601

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-07-05 for CELLEBRITY? M00516011 1601 manufactured by Boston Scientific - Spencer.

Event Text Entries

[79070789] (b)(4). Investigation results: visual analysis of the returned cellebrity cytology brush found that the yellow sheath was detached from the handle. Further evaluation noted that the proximal portion of the catheter was torn and the pull wire was kinked. There were no issues found on the handle or cannula. Functional evaluation could not be performed due to flare detachment. The complaint that the catheter was detached was confirmed. Therefore, the most probable root cause for this event is determined to be handling damage. A review of the device history record (dhr) was performed and no deviations were found. A search of the complaint database revealed that no other complaints exist for the specified lot.
Patient Sequence No: 1, Text Type: N, H10


[79070790] It was reported to boston scientific corporation that a cellebrity cytology brush was opened for use during a cytology procedure performed on (b)(6) 2017. According to the complainant, during preparation, it was noted that the catheter was detached. The procedure was completed with another cellebrity cytology brush. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be good.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2017-01938
MDR Report Key6687101
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-07-05
Date of Report2017-06-14
Date of Event2017-06-14
Date Mfgr Received2017-06-14
Device Manufacturer Date2017-02-14
Date Added to Maude2017-07-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1BOSTON SCIENTIFIC - SPENCER
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCELLEBRITY?
Generic NameBRUSH, BIOPSY, BRONCHOSCOPE (NON-RIGID)
Product CodeBTG
Date Received2017-07-05
Returned To Mfg2017-06-19
Model NumberM00516011
Catalog Number1601
Lot Number20285037
Device Expiration Date2019-02-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - SPENCER
Manufacturer Address780 BROOKSIDE DRIVE SPENCER IN 47460 US 47460


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.