FREE PSA 03289788190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-07-07 for FREE PSA 03289788190 manufactured by Roche Diagnostics.

Event Text Entries

[79526321] This event occurred in (b)(6). Unique identifier (udi)#: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[79526322] The customer stated that they received erroneous results for one patient sample tested for the elecsys free psa immunoassay (fpsa) and the elecsys total psa immunoassay (tpsa) on a cobas 8000 e 602 module (e602). The fpsa result was higher than the tpsa result. The same results occur when other blood samples of the patient are tested (the fpsa result is greater than the tpsa result). This medwatch will apply to the fpsa assay. Patient identifier (b)(6) for information related to the tpsa assay. The sample resulted as 4. 78 ng/ml for fpsa and 0. 327 ng/ml for tpsa. No adverse events were alleged to have occurred with the patient. The e602 analyzer serial number was asked for, but not provided.
Patient Sequence No: 1, Text Type: D, B5


[120550043] The patient sample was provided for investigation. The customer's results were confirmed during the investigation. The sample contained an interferent to the tpsa assay. This limitation is covered in product labeling.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2017-01402
MDR Report Key6692359
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-07-07
Date of Report2017-07-24
Date of Event2017-06-21
Date Mfgr Received2017-06-21
Date Added to Maude2017-07-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2017-07-07
Model NumberNA
Catalog Number03289788190
Lot Number20987700
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-07