MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-07-10 for COBAS 8000 C 702 MODULE C702 06473245001 manufactured by Roche Diagnostics.
[79546191]
(b)(4).
Patient Sequence No: 1, Text Type: N, H10
[79546192]
The customer stated that they received an erroneous result for one patient sample tested for ureal urea/bun (bun) on a cobas 8000 c 702 module. All other tests that were performed on the patient sample were repeated and the repeat results were the same. The sample initially resulted as 1. 5 mg/dl and this value was reported outside of the laboratory. The value was questioned by the customer since a delta check indicated that a previous sample from the patient resulted as 141 mg/dl on (b)(6) 2017. The questioned sample was repeated, resulting as 95. 6 mg/dl. The repeat result was reported outside of the laboratory and was believed to be correct. The patient was not adversely affected. The bun reagent lot number was 196836. The reagent expiration date was asked for, but not provided.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2017-01431 |
MDR Report Key | 6697965 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2017-07-10 |
Date of Report | 2017-08-13 |
Date of Event | 2017-06-23 |
Date Mfgr Received | 2017-06-24 |
Date Added to Maude | 2017-07-10 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | HITACHI HIGH TECH CORP. |
Manufacturer Street | 882 ICHIGE HITACHINAKA NA |
Manufacturer City | IBARAKI 312-8504 |
Manufacturer Country | JA |
Manufacturer Postal Code | 312-8504 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COBAS 8000 C 702 MODULE |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | CDQ |
Date Received | 2017-07-10 |
Model Number | C702 |
Catalog Number | 06473245001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Brand Name | COBAS 8000 C 702 MODULE |
Generic Name | CLINICAL CHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2017-07-10 |
Model Number | C702 |
Catalog Number | 06473245001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-07-10 |