RAPIDPOINT 500 BLOOD GAS ANALYZER 10697306

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-07-10 for RAPIDPOINT 500 BLOOD GAS ANALYZER 10697306 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[79759361] Siemens reviewed the data files. The thb results of the two patients' specimens in question appear to be valid. Co-ox diagnostic information indicates that the co-ox slide cell was stable and consistent during this testing period, showed no chronic co-ox related issues across the life of the mcart and thb recovery for the aqc samples were in range. The root-cause for the dissimilar thb results could not be determined.
Patient Sequence No: 1, Text Type: N, H10


[79759362] The customer reported discrepant elevated total hemoglobin results for two patients when run on the rp 500. There was no report of injury due to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002637618-2017-00111
MDR Report Key6699144
Date Received2017-07-10
Date of Report2017-07-10
Date of Event2017-06-21
Date Mfgr Received2017-06-21
Date Added to Maude2017-07-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEVEN ANDBERG
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone7812693655
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
Manufacturer StreetNORTHERN ROAD CHILTON INDUSTRIAL ESTATE
Manufacturer CitySUDBURY, CO102XQ
Manufacturer CountryUK
Manufacturer Postal CodeCO10 2XQ
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Sequence Number: 0

Brand NameRAPIDPOINT 500 BLOOD GAS ANALYZER
Generic NameRP 500
Product CodeGKR
Date Received2017-07-10
Catalog Number10697306
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591

Device Sequence Number: 1

Brand NameRAPIDPOINT 500 BLOOD GAS ANALYZER
Generic NameRP 500
Product CodeCHL
Date Received2017-07-10
Catalog Number10697306
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.