DISECT-PK CRABTREE 2 1/2MM FLT N1705 83J

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-07-13 for DISECT-PK CRABTREE 2 1/2MM FLT N1705 83J manufactured by Bausch + Lomb.

Event Text Entries

[79958473] The instrument was visually inspected under a microscope. There is no evidence of material defect or manufacturing error. The instrument is bent, nicked, and broken. The instrument was manufactured prior to 2009, the instrument is stamped, laser etching was introduced in 2009. The instrument has been well used, and appears to have been abused. Cannot determine why tip broke.
Patient Sequence No: 1, Text Type: N, H10


[79958474] The user facility reported during surgery the tip of the instrument broke off in the patient's ear. The doctor was able to remove the piece and continue with the surgery with no issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001920664-2017-00246
MDR Report Key6710972
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-07-13
Date of Report2017-06-21
Date of Event2017-06-15
Date Mfgr Received2017-06-21
Date Added to Maude2017-07-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JULI MOORE
Manufacturer Street3365 TREE COURT INDUSTRIAL BLV
Manufacturer CityST. LOUIS MO 63122
Manufacturer CountryUS
Manufacturer Postal63122
Manufacturer Phone636226-322
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameDISECT-PK CRABTREE 2 1/2MM FLT
Generic NameHANDHELD SURGICAL INSTRUMENT
Product CodeKAE
Date Received2017-07-13
Returned To Mfg2017-06-28
Model NumberN1705 83J
ID NumberNI
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerBAUSCH + LOMB
Manufacturer AddressROCHESTER NY 14609 US 14609

Device Sequence Number: 1

Brand NameDISECT-PK CRABTREE 2 1/2MM FLT
Generic NameHANDHELD SURGICAL INSTRUMENT
Product CodeHNR
Date Received2017-07-13
Returned To Mfg2017-06-28
Model NumberN1705 83J
ID NumberNI
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAUSCH + LOMB
Manufacturer AddressROCHESTER NY 14609 US 14609


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-13

© 2024 FDA.report