VASCULAR PROBE 7081015ES

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-07-13 for VASCULAR PROBE 7081015ES manufactured by Synovis Surgical Innovations.

Event Text Entries

[80010940] (b)(6). Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[80010941] It was reported particulate matter was observed in a vascular probe device. The unspecified particulate matter was observed in the inner pouch. The event occurred prior to patient use; therefore, there was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


[104043002] The device was received for evaluation. During visual examination, loose particulate matter (pm) was observed on the inside wall of the inner pouch. Microscopic inspection was performed and the pm was determined to be a fiber. The size was less than 0. 80mm?. Per specification, foreign material must not be larger than 0. 80 mm? ; therefore, the reported condition was not verified. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1416980-2017-05810
MDR Report Key6712945
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-07-13
Date of Report2017-08-16
Date of Event2017-04-17
Date Mfgr Received2017-08-08
Device Manufacturer Date2017-02-24
Date Added to Maude2017-07-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1SYNOVIS SURGICAL INNOVATIONS
Manufacturer Street2575 UNIVERSITY AVE W
Manufacturer CitySAINT PAUL MN 55114
Manufacturer CountryUS
Manufacturer Postal Code55114
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVASCULAR PROBE
Generic NameDILATOR, VESSEL, SURGICAL
Product CodeDWP
Date Received2017-07-13
Returned To Mfg2017-07-13
Model NumberNA
Catalog Number7081015ES
Lot NumberSP17B231215363
ID NumberNA
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSYNOVIS SURGICAL INNOVATIONS
Manufacturer AddressSAINT PAUL MN


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.