VITROS CHEMISTRY PRODUCTS DGXN SLIDES 8343386

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-07-17 for VITROS CHEMISTRY PRODUCTS DGXN SLIDES 8343386 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[80631625] The investigation determined that a lower than expected vitros dgxn result was obtained when testing a proficiency sample using a vitros 4600 chemistry system. The investigation was unable to determine the assignable cause of the event. A review of historical quality control data concluded that there is no issue with the vitros dgxn slide lot based on the dgxn qc performance. The alt and dgxn within-run marker precision results were acceptable indicating the vitros 4600 chemistry system was performing as intended. However, an ortho fe has been requested on site due to condition code u90-382 (ir wash error) that occurred one time during the dgxn precision testing. Therefore, an instrument issue cannot be entirely ruled out as a contributing factor to this event. The (b)(6) samples are from the 2017 c a survey, were initially processed (b)(6) 2017 and stored frozen. The recommended open storage is 2-8 c for 7 days. Long term storage and stability for these cap fluids is unknown, therefore, improper fluid handling (storage) cannot be ruled out as contributing to the event.
Patient Sequence No: 1, Text Type: N, H10


[80631626] The customer observed a lower than expected vitros dgxn result when testing a (b)(6) proficiency sample on a vitros 4600 chemistry system. Cap sample dgxn result 1. 7 ng/ml versus expected dgxn result 2. 81 ng/ml. Biased results of the direction and magnitude observed may lead to inappropriate physician action. The vitros dgxn result was obtained when testing a proficiency sample. There was no allegation of patient harm as a result of the event. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319809-2017-00132
MDR Report Key6718604
Date Received2017-07-17
Date of Report2017-07-17
Date of Event2017-06-19
Date Mfgr Received2017-06-19
Device Manufacturer Date2017-04-05
Date Added to Maude2017-07-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street513 TECHNOLOGY BLVD.
Manufacturer CityROCHESTER NY 14652
Manufacturer CountryUS
Manufacturer Postal Code14652
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS DGXN SLIDES
Generic NameIN VITRO DIAGNOSTIC
Product CodeLFM
Date Received2017-07-17
Catalog Number8343386
Lot Number1913-0242-0036
ID Number10758750004782
Device Expiration Date2017-09-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-17

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.