ENTACT SEPTAL STAPLER, 1-PACK 601-00100S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-07-17 for ENTACT SEPTAL STAPLER, 1-PACK 601-00100S manufactured by Arthrocare Corp..

Event Text Entries

[80299993]
Patient Sequence No: 1, Text Type: N, H10


[80299994] It was reported that when the stapler was squeezed, it cracked and all the pieces came out. It is unknown whether any of the pieces fell into the surgical site or were left in situ. It was also noted that a s&n back-up device was not available so the method/device(s) used to complete the procedure is unknown. No patient injury or other complications have been reported.
Patient Sequence No: 1, Text Type: D, B5


[115822000] Visual inspection and functional testing could not be performed because the device in question was not returned for evaluation. Thus, the complaint could not be verified and a root cause could not be determined with confidence. More than likely, upon squeezing with unintended excessive force of the device trigger this caused the handle to be become dismantled? Fall apart?. A review of the manufacturing records for the device found no deficiencies associated with the alleged event.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006524618-2017-00184
MDR Report Key6719259
Date Received2017-07-17
Date of Report2017-09-27
Date of Event2017-06-20
Date Mfgr Received2017-09-22
Date Added to Maude2017-07-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM GONZALES
Manufacturer Street7000 W. WILLIAM CANNON
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal78735
Manufacturer G1ARTHROCARE CORP.
Manufacturer Street7000 W. WILLIAM CANNON
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal Code78735
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameENTACT SEPTAL STAPLER, 1-PACK
Generic NameSEPTAL STAPLER/ABSORBABLE STAPLES
Product CodeOLL
Date Received2017-07-17
Catalog Number601-00100S
Lot Number409054
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerARTHROCARE CORP.
Manufacturer Address7000 W. WILLIAM CANNON AUSTIN TX 78735 US 78735


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-17

© 2024 FDA.report