STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903200 390-320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2017-07-17 for STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903200 390-320 manufactured by Boston Scientific - Marlborough.

Event Text Entries

[80257417] (b)(4). A visual analysis revealed that the device was returned closed. The blue sheath overran the distal stop, affecting the devices' ability to open. The blue/green shrink on the distal end of the cone was missing and was not returned. Also, the white heat shrink was found accordioned at the proximal end. Since the complaint is associated with a product which met specifications but most likely encountered anatomical or procedural factors which limited its performance, the most probable root cause for this event is operational context. The device history record (dhr) review found the device met all manufacturing specifications. A search of the complaint database revealed that no similar complaints exist for the specified lot.
Patient Sequence No: 1, Text Type: N, H10


[80257418] It was reported to boston scientific corporation that a stone cone nitinol urological retrieval coil was used in the ureter during a transurethral ureterolithotripsy (tul) procedure performed on an unknown date. According to the complainant, during the procedure, the coil of the stone cone failed to open. The procedure was completed by another of the same device. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be "good. " this event has been deemed a reportable event based on the investigation results; coil coating peeled.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2017-02020
MDR Report Key6719837
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2017-07-17
Date of Report2017-06-20
Date Mfgr Received2017-06-20
Device Manufacturer Date2016-09-06
Date Added to Maude2017-07-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1ACCELLENT ENDOSCOPY JUAREZ
Manufacturer Street31C BUTTERFIELD TRAIL
Manufacturer CityEL PASO TX 79906
Manufacturer CountryUS
Manufacturer Postal Code79906
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTONE CONE NITINOL UROLOGICAL RETRIEVAL COIL
Generic NameDISLODGER, STONE, FLEXIBLE
Product CodeFGO
Date Received2017-07-17
Returned To Mfg2017-05-17
Model NumberM0063903200
Catalog Number390-320
Lot NumberF69386
Device Expiration Date2019-07-25
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - MARLBOROUGH
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-17

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.