Event Text Entries

[80247983] Investigation conclusion: the customer's results were not replicated in-house with retention products of the reported lot. Retention products were tested with positive sample at the qc cutoff (2. 5e07 org/ml). All strips produced positive results at the read time. The product met the qc release specification. False negative results were not observed during the investigation. Manufacturing batch record review did not uncover any abnormalities. Based on the information available, there is no indication of a product deficiency and no corrective action is required.
Patient Sequence No: 1, Text Type: N, H10


[80247984] The customer reported receiving a negative strep a result using the consult strep a dipstick on five patients. For each patient, a culture method provided by (b)(6) produced a positive strep a result. As a result, each patient was delayed in receiving antibiotics for approximately 48 hours before the culture result was received.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2017-00118
MDR Report Key6720403
Date Received2017-07-17
Date of Report2017-06-19
Date Mfgr Received2017-06-19
Date Added to Maude2017-07-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCONSULT STREP A DIPSTICK 4999 50T
Generic NameSTREP A RAPID TEST STRIP
Product CodeGTY
Date Received2017-07-17
Model NumberIST-511
Lot NumberSTA6080092
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-07-17