MONOMAX VIOLET 2/0 (3) 70CM FR26 (M) B0041887

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-07-18 for MONOMAX VIOLET 2/0 (3) 70CM FR26 (M) B0041887 manufactured by B.braun Surgical Sa.

Event Text Entries

[80415405] (b)(4). Manufacturing site evaluation: samples received: 1 open pouch. Analysis and results: there is one previous complaint of this code batch. Manufactured and distributed in the market 720 units of this code batch. There are no units in stock. We have received one open sample (only the second pack is open) with the aluminum pack stuck to the paper foil of second pack. The first pack is sealed to second pack in the 4th sealing as can be seen in enclosed picture. This defect took place in the welding machine and this unit was not sorted out by the personnel involved in this process. Final conclusion: taking into account that sample received does not fulfill the specifications, it is concluded that the complaint is justified. Corrective/preventive actions: this complaint will be included in the analysis of trending, and corrective action will be open if applies.
Patient Sequence No: 1, Text Type: N, H10


[80415406] Country of complaint: (b)(6). The inner pack is sealed to the outer pack.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003639970-2017-00375
MDR Report Key6723150
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-07-18
Date of Report2017-07-18
Date Facility Aware2017-07-10
Date Mfgr Received2017-06-26
Date Added to Maude2017-07-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1B.BRAUN SURGICAL SA
Manufacturer Street121 CARRETERA DE TERRASSA
Manufacturer CityRUBI, 08191
Manufacturer CountrySP
Manufacturer Postal Code08191
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMONOMAX VIOLET 2/0 (3) 70CM FR26 (M)
Generic NameSUTURES
Product CodeNWJ
Date Received2017-07-18
Returned To Mfg2017-07-03
Model NumberB0041887
Catalog NumberB0041887
Lot Number116424
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerB.BRAUN SURGICAL SA
Manufacturer Address121 CARRETERA DE TERRASSA RUBI, 08191 SP 08191


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-07-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.