DIMENSION? RF420 SMN 10444904

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-07-19 for DIMENSION? RF420 SMN 10444904 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[80412922] The cause of the discordant elevated mmb result is unknown. Customer care center (ccc) analysis of the data shows imprecision as the sample was rerun. Values ranged from just below the assay upper assay range to values greatly above the range. Cki and tni were run on same patient and no issues were reported with the patient results. Quality control was within laboratory range on the date of testing. Siemens is investigating the incident.
Patient Sequence No: 1, Text Type: N, H10


[80412923] A discordant elevated mass creatine kinase mb isoenzyme (mmb) result was obtained on a patient sample on the dimension exl system. The repeated result >600 ng/ml was reported to the physician and not questioned. The same sample was auto-diluted and elevated results were obtained. Manual dilution of the sample (2x and 5x) obtained lower results. There are no known reports of patient intervention or adverse health consequences due to the discordant elevated mmb result.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2017-00569
MDR Report Key6724346
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-07-19
Date of Report2017-08-16
Date of Event2017-06-20
Date Mfgr Received2017-07-25
Device Manufacturer Date2017-04-03
Date Added to Maude2017-07-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal Code197146101
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION?
Generic NameDIMENSION? MMB MASS CREATINE KINASE MB ISOENZYME FLEX? REAGENT CARTRIDGE
Product CodeJHY
Date Received2017-07-19
Catalog NumberRF420 SMN 10444904
Lot NumberFB8093
Device Expiration Date2018-04-03
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.