COBAS 6000 E 601 MODULE E601 04745922001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-07-20 for COBAS 6000 E 601 MODULE E601 04745922001 manufactured by Roche Diagnostics.

Event Text Entries

[80789470] Unique identifier (udi)#: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[80789491] The customer received a questionable elecsys ferritin result for one patient sample. The initial result was 1. 02 ng/ml and the repeat result was 39. 94 ng/ml. The initial result was reported outside the laboratory. The repeat result was believed to be correct. The patient was not adversely affected. The reagent lot number was 20990805 with an expiration date of 5/31/2018. The field service representative flushed the reagent probe, checked and cleaned the bead mixer and sipper probe, and visually inspected measuring cell. The customer ran calibration and qc. No abnormal trends were identified for this issue. The customer has had no other complaints of this nature on a like analyzer in the past 12 months.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-01506
MDR Report Key6728024
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-07-20
Date of Report2017-07-20
Date of Event2017-06-29
Date Mfgr Received2017-06-30
Date Added to Maude2017-07-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 6000 E 601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJMG
Date Received2017-07-20
Model NumberE601
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 6000 E 601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-07-20
Model NumberE601
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-20