STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903100 390-310

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2017-07-21 for STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903100 390-310 manufactured by Boston Scientific - Marlborough.

Event Text Entries

[80620175] (b)(4). Although the suspect device has been received, the evaluation has not been completed. Therefore, the cause of the reported malfunction has not been determined. Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10


[80620176] It was reported to boston scientific corporation that a stone cone nitinol urological retrieval coil was used in the ureter on a procedure performed on (b)(6) 2017. According to the complainant, during the procedure, the stone cone retrieval device failed to retract into the sheath. The retrieval device was removed together with the scope from the patient's body. However, when the device was removed it was found the tip broke and detached. A grasping forceps was used to retrieved the detached tip. Fluoroscopy confirmed that no device fragments are left inside the patient. Additionally, it was reported that a laser was used during the procedure but it did not damage the retrieval device. There were no patient complications reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2017-02080
MDR Report Key6731037
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2017-07-21
Date of Report2017-06-26
Date of Event2017-06-20
Date Mfgr Received2017-07-28
Device Manufacturer Date2016-08-08
Date Added to Maude2017-07-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1ACCELLENT ENDOSCOPY JUAREZ
Manufacturer Street31C BUTTERFIELD TRAIL
Manufacturer CityEL PASO TX 79906
Manufacturer CountryUS
Manufacturer Postal Code79906
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTONE CONE NITINOL UROLOGICAL RETRIEVAL COIL
Generic NameDISLODGER, STONE, FLEXIBLE
Product CodeFGO
Date Received2017-07-21
Returned To Mfg2017-06-28
Model NumberM0063903100
Catalog Number390-310
Lot NumberF52633
Device Expiration Date2019-06-09
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - MARLBOROUGH
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-07-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.