DIMENSION? DIMENSION EXL WITH LM SMN 10472176

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-07-25 for DIMENSION? DIMENSION EXL WITH LM SMN 10472176 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[80934875] The cause for the discordant depressed actm result is instrument malfunction. A siemens customer service engineer was dispatched to the site and performed service including maintenance on the optical system. They also checked cuvette manufacturing, cleaned drains, and checked syringes. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[80934876] A discordant falsely depressed acetaminophen (actm) result was obtained on the dimension exl instrument. The result was flagged below assay range and was not reported to the physician. The same sample was repeated on the same instrument and on an alternate dimension exl instrument and a higher result was obtained and reported. There are no known reports of patient intervention or adverse health consequences due to the discordant depressed actm result.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2017-00605
MDR Report Key6739668
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-07-25
Date of Report2017-07-25
Date of Event2017-06-07
Date Mfgr Received2017-06-29
Device Manufacturer Date2011-11-21
Date Added to Maude2017-07-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC- BROOKFIELD
Manufacturer StreetREGISTRATION NUMBER 1226181 101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameDIMENSION?
Generic NameDIMENSION? EXL WITH LM INTEGRATED CHEMISTRY SYSTEM
Product CodeLDP
Date Received2017-07-25
Model NumberDIMENSION EXL WITH LM
Catalog NumberSMN 10472176
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101

Device Sequence Number: 1

Brand NameDIMENSION?
Generic NameDIMENSION? EXL WITH LM INTEGRATED CHEMISTRY SYSTEM
Product CodeJJE
Date Received2017-07-25
Model NumberDIMENSION EXL WITH LM
Catalog NumberSMN 10472176
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.