TENDER CARE HYDROGEL PADS 87123NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2017-07-25 for TENDER CARE HYDROGEL PADS 87123NA manufactured by Medela Llc.

Event Text Entries

[80987313] A replacement box of hydrogel pads were sent to the customer so she could return them to a store for a refund. On (b)(6) 2017, a medela clinician followed up with the customer and she reported that she went to urgent care and received a steroid shot and also went to her doctor and was prescribed diprolene cream. She has not used the pads since the one time on (b)(6) 2017. She indicated that the rash was clearing up, but as of (b)(6) 2017, there was still redness outlining the pads. The product involved in the complaint has not been returned for evaluation/investigation at this time. Reported issues of rash were investigated under in (b)(4), which determined the probable root causes to be (1) customers with glycerin allergies experiencing aggressive immune reactions due to the fact that glycerin is an integral ingredient in the hydrogel pads or (2) pathogens or allergens that come in contact with exposed hydrogel pads which would subsequently be introduced to the user? S breast. The customer filed a voluntary medwatch report under mw5070852.
Patient Sequence No: 1, Text Type: N, H10


[80987314] On (b)(6) 2017, the customer alleged that she used the tender care hydrogel pads on (b)(6) 2017 and was apparently allergic to them, as they caused her skin to become red and she developed a rash covering the area where the pad was.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419937-2017-00206
MDR Report Key6742009
Report SourceCONSUMER
Date Received2017-07-25
Date of Report2017-07-25
Date of Event2017-07-03
Date Facility Aware2017-07-11
Date Mfgr Received2017-07-11
Date Added to Maude2017-07-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactROBERT SOKOLOWSKI
Manufacturer Street1101 CORPORATE DRIVE
Manufacturer CityMCHENRY IL 60050
Manufacturer CountryUS
Manufacturer Postal60050
Manufacturer G1MEDELA LLC
Manufacturer Street1101 CORPORATE DRIVE
Manufacturer CityMCHENRY IL 60050
Manufacturer CountryUS
Manufacturer Postal Code60050
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTENDER CARE HYDROGEL PADS
Generic NameNURSING PAD, HYDROGEL
Product CodeNTC
Date Received2017-07-25
Model Number87123NA
Catalog Number87123NA
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDELA LLC
Manufacturer Address1101 CORPORATE DRIVE MCHENRY IL


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-07-25