LIGHT FANTASTIC II TRACK MOUNTED DENTAL LIGHT HBT LFTII

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-07-25 for LIGHT FANTASTIC II TRACK MOUNTED DENTAL LIGHT HBT LFTII manufactured by Pelton & Crane.

Event Text Entries

[81245357] Upon evaluation by the distributor, it was determined the roll pins and set screws were not installed by the distributor during installation. The roll pins and set screws prevents the light from unscrewing from the trolley after installation. The distributor service technician ordered the roll pins and set screws and properly installed them on the dental light. The dental light was manufactured in may 1979 and is over 38 years old and past the expected life of the device.
Patient Sequence No: 1, Text Type: N, H10


[81245358] It was reported that a pelton & crane dental light fell from the track assembly and struck the patient on the head causing a small laceration above the patients eye. General first aid was provided only. There were no serious injuries reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017522-2017-00037
MDR Report Key6742387
Date Received2017-07-25
Date of Report2017-06-28
Date of Event2017-06-28
Date Mfgr Received2017-06-28
Device Manufacturer Date1979-05-11
Date Added to Maude2017-07-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. FRANK RAY
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1PELTON & CRANE
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal Code28273
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameLIGHT FANTASTIC II TRACK MOUNTED DENTAL LIGHT
Generic NameDENTAL LIGHT
Product CodeEAZ
Date Received2017-07-25
Model NumberHBT
Catalog NumberLFTII
OperatorDENTAL HYGIENIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPELTON & CRANE
Manufacturer Address11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2017-07-25

© 2026 FDA.report